ATELLICA IM BR 27.29 (BR)
Received Jun 18, 2026 · Event occurred May 7, 2026
Report 1219913-2026-00117 · MDR key 25546257
Device
Generic name
System, Test, Immunological, Antigen, Tumor
Manufacturer
Siemens Healthcare Diagnostics, Inc.Model number
N/ACatalog number
10995477
Lot number
87285275
Product problems
- High Test Results
Patient
NA · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE CUSTOMER REPORTS OBSERVATION OF ELEVATED ATELLICA IM BR 27.29 (BR) RESULTS WHICH WERE DISCORDANT RELATIVE TO REPEAT TESTING. SIEMENS IS INVESTIGATING. NOTE: SIMILAR EVENTS ASSOCIATED WITH OTHER OBSERVATION DATES ARE REPORTED SEPARATELY.
Description of Event or Problem
THE CUSTOMER REPORTS OBSERVATION OF ELEVATED QUALITY CONTROL (QC) AND DISCORDANT PATIENT RESULTS WHEN RUNNING ATELLICA IM BR 27.29 (BR) LOT 275 WHEN REMAINING REAGENT VOLUME IS LOW. THE CUSTOMER COMPLAINED OF ELEVATED QC RESULTS PRODUCED WHEN FEWER TESTS (<30) ARE REMAINING IN THE BR 27.29 REAGENT PACK. THIS ISSUE WAS SEEN ACROSS ALL THREE ATELLICA IM ANALYZERS IN THIS LAB. ADDITIONAL TESTING USING DILUTED CONTROL MATERIAL DEMONSTRATED A SIMILAR EFFECT. ON 09-JUN-2026, THE CUSTOMER PROVIDED DATA INDICATING THAT DISCORDANT PATIENT RESULTS HAD BEEN OBSERVED IN ASSAY RUNS WHERE QC HAD FAILED. DATA WERE PROVIDED FOR SEVERAL MONTHS OF OBSERVATIONS, WHERE QC RESULTS WERE OUTSIDE EXPECTED RANGES, AND PATIENT SAMPLES WHICH HAD BEEN TESTED SINCE THE LAST PASSING QC WERE RE-TESTED. RESULTS FOR PATIENT SAMPLES WHICH HAD BEEN SUBJECTED TO REPEAT TESTING WERE PROVIDED. DISCORDANCE WAS IDENTIFIED BETWEEN INITIAL AND REPEAT RESULTS FOR SIX SAMPLES TESTED ON THIS DATE (B)(6), WHERE THE INITIAL RESULTS WERE IDENTIFIED AS FALSELY ELEVATED. THERE ARE NO ALLEGATIONS OF ANY ADVERSE PATIENT CONSEQUENCES IN ASSOCIATION WITH THE OBSERVED DISCORDANCE.