Received Jun 17, 2026 · Event occurred May 20, 2026

Report 1824619-2026-00017 · MDR key 25535973

Device

Generic name

Circlamp W/1.3 Cm Bell (st)

Catalog number

330CR

Lot number

2026020301

Product problems

  • Appropriate Device Problem Term/Code Not Available

Patient

NA · Unknown

  • Genital Bleeding

Narrative

Description of Event or Problem

ACCORDING TO THE FACILITY, THEY HAVE EXPERIENCED FOUR CASES OF SIGNIFICANT BLEEDING FOLLOWING CIRCUMCISION PROCEDURES. CLINICAL STAFF REPORT THAT THESE DISPOSABLE CLAMPS DO NOT SEEM TO TIGHTEN SUFFICIENTLY, RESULTING IN VISIBLE BLOOD LEAKAGE.

Additional Manufacturer Narrative

ACCORDING TO THE FACILITY, THEY HAVE EXPERIENCED FOUR CASES OF SIGNIFICANT BLEEDING FOLLOWING CIRCUMCISION PROCEDURES. CLINICAL STAFF REPORT THAT THESE DISPOSABLE CLAMPS DO NOT SEEM TO TIGHTEN SUFFICIENTLY, RESULTING IN VISIBLE BLOOD LEAKAGE. OUR PEDIATRICIAN COMPARED THE DISPOSABLE CLAMPS TO OUR NON-DISPOSABLE VERSIONS AND NOTED A SIGNIFICANT DIFFERENCE IN SIZING, WITH THE DISPOSABLES BEING LARGER. ADDITIONALLY, WHEN ATTEMPTING TO TIGHTEN THE DISPOSABLE CLAMP, IT CONTINUES TO SPIN RATHER THAN REACHING A SECURE TENSION LIKE THE NON-DISPOSABLE VERSION. NO ADDITIONAL INFORMATION IS AVAILABLE. IT HAS BEEN DETERMINED THAT THOUGH THE EVENT DID NOT INVOLVE A DEATH OR SERIOUS INJURY, RECURRENCE COULD CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY. BASED ON THE INFORMATION REVIEWED, THIS IS A REPORTABLE EVENT AS DEFINED UNDER 21 CFR 803.3.