inforMED
InjuryJCW

AMS 700 LGX MS PUMP

Received Jun 17, 2026 · Event occurred Jun 4, 2026

Report 2124215-2026-32333 · MDR key 25535461

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404251

Catalog number

72404251

Lot number

1000337800

Product problems

  • Fluid/Blood Leak
  • Mechanical Problem
  • Material Puncture/Hole

Patient

72 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MODEL NUMBER/CATALOG NUMBER: 72404155. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1000451610. MODEL/CATALOG DESCRIPTION: RESERVOIR 65 ML PC/IZ. UNIQUE IDENTIFIER (UDI) #: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT WORKING PROPERLY. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH THE DEVICE WAS EXPLANTED AND REPLACED. UPON EXPLANT THE PHYSICIAN IDENTIFIED A TEAR IN ONE OF THE CYLINDERS AND LOSS OF FLUID AS THERE WAS NO FLUID IN THE SYSTEM. THERE WERE NO PATIENT COMPLICATIONS REPORTED.