AMS 700 LGX MS PUMP
Received Jun 17, 2026 · Event occurred Jan 1, 2026
Report 2124215-2026-32322 · MDR key 25535438
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
72404252Catalog number
72404252
Lot number
0175669004
Product problems
- Fluid/Blood Leak
- Inflation Problem
- Material Puncture/Hole
- Connection Problem
Patient
57 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MODEL NUMBER/CATALOG NUMBER: 72404156, SERIAL NUMBER: N/A, BATCH/LOT NUMBER: 0178950001, MODEL/CATALOG DESCRIPTION: RESERVOIR 100 ML PC/IZ, UNIQUE IDENTIFIER (UDI) #: (B)(4). THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. BASED ON THE INFORMATION AVAILABLE THE CAUSE THAT CONTRIBUTED TO THE REPORTED EVENT CANNOT BE ESTABLISHED. THEREFORE, A CONCLUSION CODE OF CAUSE NOT ESTABLISHED WAS ASSIGNED TO THIS INVESTIGATION.
Description of Event or Problem
IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) EXHIBITED INFLATION ISSUES. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH IT WAS IDENTIFIED A FLUID LOSS DUE TO A TUBING DISCONNECTION AND BOTH CYLINDERS RUPTURED REPORTED FROM WEAR AND TEAR. THE IPP WAS EXPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.