inforMED
InjuryFAD

TUMOR STENT DOUBLE LOOP URETERAL STENT KIT

Received Jun 16, 2026 · Event occurred May 1, 2025

Report 9610711-2026-00189 · MDR key 25524069

Device

Generic name

Urological Surgical Procedure Kit, Non-medicated, Single Use

Manufacturer

Coloplast A/s

Catalog number

BCCG74

Lot number

10669885_BCCG744002

Product problems

  • Device Stenosis

Patient

NA · Unknown

  • Renal Failure

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION IS NOT YET COMPLETE, A FOLLOW UP REPORT WILL BE SUBMITTED ON COMPLETION OF THE INVESTIGATION. THE ADDITIONAL DEVICES REFERENCED IN B5 ARE CAPTURED UNDER SEPARATE REPORTS.

Description of Event or Problem

ACCORDING TO THE AVAILABLE INFORMATION, CLOGGED STENTS REQUIRING EARLY REPLACEMENT. STENT HAD TOTAL OBSTRUCTION. THE STENTS WERE PLACED ON A SCHEDULED BASIS BUT CHANGED ON AN EMERGENCY BASIS. 1 TO MULTIPLE REHOSPITALIZATIONS PER PATIENT DUE TO ACUTE RENAL FAILURE. EARLY REPLACEMENT OF STENTS UNDER GENERAL OR LOCAL ANESTHESIA IN CANCER PATIENTS. AVERAGE TIME BEFORE OBSTRUCTION WAS OBSERVED WAS 89 DAYS BETWEEN THE PLACEMENT OF THE FIRST CATHETER AND ITS EARLY REPLACEMENT. IMPLANTATIONS WERE MOSTLY FOR BLADDER TUMORS, MOSTLY EXTERNAL COMPRESSION. DEVICE WAS EXPOSED TO UNUSUAL CONDITIONS AND FREE OF ANY VISIBLE DEFECTS UPON OPENING.