ALINITY C CARBON DIOXIDE REAGENT KIT
Received Jun 16, 2026 · Event occurred May 18, 2026
Report 3002809144-2026-00160 · MDR key 25520971
Device
Generic name
Enzymatic, Carbon-dioxide
Manufacturer
Abbott GmbhCatalog number
07P72-20
Lot number
69531UQ01
Product problems
- Low Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY DEPRESSED ALINITY C CARBON DIOXIDE RESULTS ON MULTIPLE PATIENTS AND MULTIPLE ANALYZERS WHEN PROCESSED ON THE ALINITY C PROCESSING MODULE. THE FOLLOWING EXAMPLES WERE PROVIDED. (CUSTOMER¿S REFERENCE RANGE 22 - 32 MMOL/L): PAT B, INITIAL RESULT = 10 MMOL/L, REPEAT RESULTS = 25 MMOL/L; PAT C, INITIAL RESULT = 11 MMOL/L, REPEAT RESULTS = 26 MMOL/L; PAT D, INITIAL RESULT = 8 MMOL/L, REPEAT RESULTS = 22 MMOL/L; PAT E, INITIAL RESULT = 8 MMOL/L, REPEAT RESULTS = 22 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.