inforMED
MalfunctionIOQ

TOTALCARE BARIATRIC PLUS BED

Received Jun 15, 2026 · Event occurred May 20, 2026

Report 1824206-2026-01348 · MDR key 25510745

Device

Generic name

Bed, Flotation Therapy, Powered

Model number

P1840R

Catalog number

P1840RE200

Product problems

  • Positioning Failure

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE BAXTER TECHNICIAN FOUND THE HYDRAULIC VALVE NEEDED TO BE REPLACED. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN (B)(6) 2025. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. PER THE BAXTER SERVICE MANUAL THE TOTALCARE BED SYSTEM REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM A SEMI-ANNUAL PREVENTATIVE MAINTENANCE. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. TEST THE HEAD SECTION SLIDE PIVOTS, SELF-LUBRICATING BEARING, AND SLIDER PIVOT EXTRUSIONS. LOOK FOR DAMAGE AND MAKE SURE THE HEAD SECTION OPERATES CORRECTLY. REPLACE COMPONENTS AS NECESSARY. THE TECHNICIAN REPLACED THE SOLENOID VALVE TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem

BAXTER RECEIVED A REPORT THAT TC BARIATRIC PLUS W/ AIR HAD HEAD OF BED NOT RAISING. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.