Device

Generic name

Integration System

Model number

HV1000

Catalog number

XEH1BASE

Lot number

(10)3.0.0

Product problems

  • Display or Visual Feedback Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE HV1000 INTEGRATION SYSTEM AND FOUND THAT THE SWITCH BOARD COMPONENT REQUIRED A REBOOT. THE TECHNICIAN REBOOTED THE SWITCH BOARD COMPONENT, TESTED THE FUNCTION AND OPERATION OF THE UNIT, CONFIRMED IT TO BE OPERATING TO SPECIFICATION, AND RETURNED THE UNIT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT THE TOUCH PANEL CONNECTED TO THEIR HV1000 INTEGRATION SYSTEM WAS FROZEN. NO REPORT OF INJURY.