MalfunctionPZP

HYDROS ROBOTIC SYSTEM

Received Jun 15, 2026 · Event occurred May 22, 2026

Report 3012977056-2026-00124 · MDR key 25508666

Device

Generic name

Fluid Jet Removal System

Catalog number

HY1000T

Product problems

  • Visual Prompts will not Clear

Patient

NA · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ROOT CAUSE OF THE REPORTED EVENT HAS NOT YET BEEN ESTABLISHED. INVESTIGATION BY MANUFACTURER IS CURRENTLY IN-PROCESS. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

Description of Event or Problem

A MALE PATIENT UNDERWENT AQUABLATION THERAPY TO TREAT SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT DURING THE FIRST TREATMENT PASS, THE HYDROS ROBOTIC SYSTEM DISPLAYED ERROR E251 (Z ENCODER SPEED WAS LESS THAN 80% OF THE DESIRED SPEED FOR OVER 10MS) AND ERROR E206 (MOTORPACK ERROR). DESPITE TROUBLESHOOTING ATTEMPTS, THE ISSUES PERSISTED. THE REPORTED EVENT RESULTED IN THE PROCEDURE BEING ABORTED WHILE THE PATIENT WAS UNDER ANESTHESIA. THERE WERE NO ADVERSE HEALTH CONSEQUENCES FOR THE PATIENT AS A RESULT OF THIS EVENT.