HYDROS ROBOTIC SYSTEM
Received Jun 15, 2026 · Event occurred May 22, 2026
Report 3012977056-2026-00124 · MDR key 25508666
Device
Generic name
Fluid Jet Removal System
Manufacturer
Procept Biorobotics CorporationCatalog number
HY1000T
Product problems
- Visual Prompts will not Clear
Patient
NA · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ROOT CAUSE OF THE REPORTED EVENT HAS NOT YET BEEN ESTABLISHED. INVESTIGATION BY MANUFACTURER IS CURRENTLY IN-PROCESS. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
Description of Event or Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY TO TREAT SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT DURING THE FIRST TREATMENT PASS, THE HYDROS ROBOTIC SYSTEM DISPLAYED ERROR E251 (Z ENCODER SPEED WAS LESS THAN 80% OF THE DESIRED SPEED FOR OVER 10MS) AND ERROR E206 (MOTORPACK ERROR). DESPITE TROUBLESHOOTING ATTEMPTS, THE ISSUES PERSISTED. THE REPORTED EVENT RESULTED IN THE PROCEDURE BEING ABORTED WHILE THE PATIENT WAS UNDER ANESTHESIA. THERE WERE NO ADVERSE HEALTH CONSEQUENCES FOR THE PATIENT AS A RESULT OF THIS EVENT.