inforMED
InjuryLIO

APTIMA MULTITEST SPECIMEN COLLECTION KIT

Received Jun 15, 2026 · Event occurred May 28, 2026

Report 2024800-2026-00033 · MDR key 25506169

Device

Generic name

Device, Specimen Collection

Manufacturer

Hologic, Inc.

Catalog number

ASY-08848

Lot number

931935

Product problems

  • Material Separation

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ON (B)(6), 2026, A CUSTOMER REPORTED TO HOLOGIC THAT THE TIP OF MULTIPLE SWABS, INCLUDED IN THE APTIMA MULTITEST SPECIMEN COLLECTION KIT (LN 931935) DETACHED FROM THE SWABS SHAFT AND REMAINED INSIDE THE PATIENTS DURING SPECIMEN SELF-COLLECTION. THE CUSTOMER DID NOT SPECIFY HOW THE SWAB TIP WAS REMOVED FROM THE PATIENT. NO FURTHER INFORMATION ON THE PATIENT¿S STATUS WAS PROVIDED TO HOLOGIC. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE OUTCOMES RELATED TO THIS EVENT.

Additional Manufacturer Narrative

A REVIEW OF THE PART HISTORY FILE ASSOCIATED WITH THE MULTITEST SWABS INCLUDED IN THE APTIMA MULTITEST SWAB KIT (LN 931935) CONFIRMED THAT THE SWABS SUCCESSFULLY PASSED INSPECTION SPECIFICATIONS. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE OUTCOMES RELATED TO THIS ISSUE. A SUPPLIER CORRECTIVE ACTION REQUEST (SCAR) HAS BEEN ISSUED TO THE SUPPLIER OF THE SWABS (PURITAN MEDICAL PRODUCTS COMPANY, LLC).