inforMED
InjuryHCG

TARGET 3D 6MM X 15CM

Received Jun 15, 2026 · Event occurred May 20, 2026

Report 3008881809-2026-00364 · MDR key 25505618

Device

Generic name

Device, Neurovascular Embolization

Catalog number

M0036156150

Lot number

24695651

Product problems

  • Migration

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT DURING DOUBLE-CATHETER FRAMING WITH TWO COILS, THE SUBJECT COIL BECAME STRETCHED AND MIGRATED INTO THE ACA (ANTERIOR CEREBRAL ARTERY); HOWEVER, PARENT VESSEL PATENCY WAS PRESERVED WITH STENT DEPLOYMENT. ADDITIONAL COILS WERE PLACED, AND THE PROCEDURE WAS COMPLETED USING THE SAME MICROCATHETERS. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.