SYNDESMOSIS TIGHTROPE PRO IMPLANT, SS

Received Jun 15, 2026 · Event occurred May 26, 2026

Report 1220246-2026-03403 · MDR key 25503221

Device

Generic name

Washer, Bolt, Nut

Manufacturer

Arthrex, Inc.

Catalog number

AR-9925SS

Lot number

15611231

Product problems

  • Detachment of Device or Device Component

Patient

NA · Unknown

  • Insufficient Information

Narrative

Description of Event or Problem

ON 28-MAY-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO AR-9925SS SYNDESMOSIS TIGHTROPE PRO DEVICES EXPERIENCED AN ISSUE IN WHICH, DURING INSERTION, THE BUTTONS DETACHED FROM THE INSERTER WITH LITTLE TO NO RESISTANCE. THE EVENTS WERE DISCOVERED DURING AN ANKLE ORIF PROCEDURE PERFORMED ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PIECES BROKE OFF INSIDE THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 03-JUN-2026: THE ANCHOR WAS REMOVED, AND NO COMPONENTS REMAINED IN THE PATIENT. NO DEVICE BREAKAGE OCCURRED WITHIN THE PATIENT. AN X-RAY WAS PERFORMED TO CONFIRM THAT NO FRAGMENTS WERE RETAINED. THE PROCEDURE WAS COMPLETED USING AN AR-8925SS TIGHTROPE XP, WHICH FUNCTIONED AS INTENDED. THE CASE EXPERIENCED AN APPROXIMATE 15-MINUTE DELAY WHILE TROUBLESHOOTING THE INITIAL SYSTEM. IT IS UNKNOWN WHETHER ADDITIONAL ANESTHESIA WAS ADMINISTERED.

Additional Manufacturer Narrative

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.