inforMED
InjuryLIO

APTIMA MULTITEST SPECIMEN COLLECTION KIT

Received Jun 15, 2026 · Event occurred May 20, 2026

Report 2024800-2026-00031 · MDR key 25502992

Device

Generic name

Device, Specimen Collection

Manufacturer

Hologic, Inc.

Catalog number

ASY-08848

Lot number

925009, 931935

Product problems

  • Material Separation

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A REVIEW OF THE PART HISTORY FILE ASSOCIATED WITH THE MULTITEST SWABS INCLUDED IN THE APTIMA MULTITEST SWAB KIT (LNS 931935 AND 925009) CONFIRMED THAT THE SWABS SUCCESSFULLY PASSED INSPECTION SPECIFICATIONS. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE OUTCOMES RELATED TO THIS ISSUE. A SUPPLIER CORRECTIVE ACTION REQUEST (SCAR) HAS BEEN ISSUED TO THE SUPPLIER OF THE SWABS (B)(4).

Description of Event or Problem

ON (B)(6) 2026, A CUSTOMER REPORTED TO HOLOGIC THAT THE TIP OF MULTIPLE SWABS, INCLUDED IN THE APTIMA MULTITEST SPECIMEN COLLECTION KIT (LNS 931935 AND 925009) DETACHED FROM THE SWABS SHAFT AND REMAINED INSIDE THE PATIENTS DURING SPECIMEN COLLECTION. THE CUSTOMER CONFIRMED THAT MEDICAL ASSISTANCE WAS REQUIRED TO REMOVE THE SWAB TIPS FROM THE PATIENTS. NO FURTHER INFORMATION ON THE PATIENTS¿ STATUS WAS PROVIDED TO HOLOGIC. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE OUTCOMES RELATED TO THIS EVENT.