Received Jun 15, 2026 · Event occurred Jun 7, 2026
Report 25501493 · MDR key 25501493
Device
Generic name
Electrode, Electrosurgical, Active, Urological
Manufacturer
Karl Storz Se & Co KgModel number
27040GP130-SCatalog number
27040GP130-S
Lot number
43LE4523
Patient
66 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
FROM STAFF: WE WERE DOING A CLOT EVACUATION/BLADDER TUMOR RESECTION CASE WITH THE STORZ RESECTOSCOPE USING THE BIPOLAR LOOP. NORMAL PRE-SET SETTINGS WERE USED DURING THE CASE. WHILE BEING USED, THERE WAS A SUDDEN POP, SPARK AND CLOUD OF SMOKE THAT CAME FROM THE CORD OF THE SCOPE. SCOPE WAS PULLED OUT AND TAKEN OFF THE FIELD, PIECES SET ASIDE TO BE EVALUATED BY RISK. THERE WAS NO NOTED HARM TO THE PATIENT. DURING SET UP, THE CORD WAS NOT NOTED TO HAVE ANY IMPERFECTIONS, HOWEVER UPON INSPECTION AFTER THE EVENT TOOK PLACE, IT APPEARS ONE WIRE IS STICKING OUT OF THE CORD AND NOT COATED.