HV1000
Received Jun 12, 2026 · Event occurred Aug 26, 2025
Report 3008776287-2026-00315 · MDR key 25491063
Device
Generic name
Integration System
Manufacturer
Black Diamond Video, Inc.Model number
HV1000Catalog number
XEH1BASE
Lot number
(10)2.2.1
Product problems
- Power Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE HV1000 INTEGRATION SYSTEM AND FOUND THAT SWITCHBOARD REQUIRED A REBOOT. THE TECHNICIAN REBOOTED THE SWITCHBOARD COMPONENT, TESTED THE FUNCTION AND OPERATION OF THE UNIT, CONFIRMED IT TO BE OPERATING TO SPECIFICATION, AND RETURNED THE UNIT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT THEIR HV1000 INTEGRATION SYSTEM WOULD NOT POWER ON. NO REPORT OF INJURY.