MalfunctionCFN

IMMAGE IMMUNOCHEMISTRY SYSTEM

Received Apr 25, 2012 · Event occurred Apr 5, 2012

Report 2050012-2012-01037 · MDR key 2548783

Device

Generic name

Nephelometer, For Clinical Use

Model number

N/A

Catalog number

A15445

Product problems

  • Overfill

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

ON (B)(6) 2012 CUSTOMER REPORTED THAT THEY HAD REPLACED THE FILTER ON SAMPLE PROBE WASH STATION, ON THE IMMAGE 800 INSTRUMENT, AND IT OVERFLOWED. NO INJURIES WERE REPORTED AND NO ERRONEOUS PATIENT RESULTS WERE GENERATED. FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND THE FILTER WAS PLUGGED. FSE REPLACED THE FILTER AND THIS RESOLVED THE ISSUE. NO FURTHER PROBLEMS WERE REPORTED.

Additional Manufacturer Narrative

(B)(4).