inforMED
MalfunctionCJY

ALINITY C CALCIUM2

Received Jun 12, 2026 · Event occurred Jun 5, 2026

Report 3005094123-2026-00282 · MDR key 25486142

Device

Generic name

Azo Dye, Calcium

Catalog number

04T87-20

Lot number

82015UD00

Product problems

  • High Test Results

Patient

37 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.

Description of Event or Problem

THE CUSTOMER REPORTED FALSELY ELEVATED CALCIUM RESULTS GENERATED FROM ALINITY C PROCESSING MODULE. THE CUSTOMER PROVIDED THE FOLLOWING DATA WHICH OCCURRED ON (B)(6) 2026: THE CALCIUM RESULT ON SN: (B)(6) WAS 3.77 MMOL/L. THE CALCIUM RESULT ON SN: (B)(6) WAS 4.12 MMOL/L. THE CALCIUM RESULT ON SN: (B)(6) WAS 2.96 MMOL/L. THE CALCIUM RESULT ON SN: (B)(6) WAS 2.94 MMOL/L. THE CALCIUM RESULT ON SN: (B)(6) WAS 2.41 MMOL/L. THE CUSTOMER REPORTED OUT THE 2.41 RESULT. CUSTOMER NORMAL RANGE IS 2.20-2.60 MMOL/L. PATIENT INFORMATION: 37-YEAR-OLD FEMALE. PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.