G30-01 VIS-RX MICRO-IMAGING CATHETER

Received Jun 12, 2026 · Event occurred Apr 19, 2023

Report 3011578544-2026-00056 · MDR key 25485275

Device

Generic name

System, Imaging, Optical Coheren

Manufacturer

Gentuity, Llc

Model number

G10-01

Catalog number

G10-01

Lot number

22J0101

Product problems

  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

BOTH DOCTORS (B)(6) AT (B)(6) SAID THAT THEY ARE CONCERNED ABOUT THE ACCURACY OF THE LENGTH OF OUR SYSTEM. DR (B)(6) SAID THAT "THEY (THE GROUP) ARE ALL TALKING ABOUT IT, AND THEY DON'T BELIEVE IT." WHEN I DID A CASE WITH DR (B)(6) HE PLACED A 23 MM ABBOTT STENT, AND WHEN I MEASURED IT, THE SYSTEM MEASURED 27 MM. SO I WENT BACK AND LOOKED AT DOCTOR (B)(6) CASE FROM THE DAY PRIOR, AND THAT WAS A 26 MM BIOTRONIC STENT, AND IT MEASURED 26 MM EXACTLY. (I DID NOT GET THE OPPORTUNITY TO SHARE THIS WITH THE PHYSICIANS). I DID DOWNLOAD BOTH CASES, TRIED TO UPLOAD, BUT INTERNET WAS POOR, SO UNABLE TO, WILL RE-TRY WHEN I RETURN HOME ON (B)(6) 2023.

Additional Manufacturer Narrative

LITERATURE REVIEW DETERMINED THAT MOTION ARTIFACT CAN CAUSE MEASUREMENTS TO VARY BY A FEW MILLIMETERS. NO INDICATION OF A PRODUCT MALFUNCTION WAS FOUND. MOTION ARTIFACT COULD EXPLAIN THE DISCREPANCY BETWEEN THE MEASUREMENT BY THE GENTUITY SYSTEM AND THE DETERMINED STENT LENGTH FOR TREATMENT.