ESOFLIP
Received Jun 12, 2026 · Event occurred Mar 30, 2026
Report 9681384-2026-00659 · MDR key 25485051
Device
Generic name
Esophageal Dilator Balloon With Or Without Ele
Manufacturer
Covidien Ag Juarez (mmj)Model number
ES-330Catalog number
ES-330
Lot number
24H1470JZ
Product problems
- Contamination /Decontamination Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS NOT RETURNED, BUT THREE PHOTOS WERE AVAILABLE FOR EVALUATION. INSPECTION OF THE IMAGES NOTED BLACK PARTICULATE MATTER SUSPENDED IN THE SALINE SOLUTION. BIOLOGICAL CONTAMINATION HAS BEEN IDENTIFIED IN PRIOR SYRINGES CONTAINING PARTICULATE CONTAMINATION. IT WAS REPORTED THAT A FOREIGN OBJECT WAS FOUND IN THE PACKAGE OR PRODUCT. THE REPORTED ISSUE WAS CONFIRMED. THE MOST LIKELY CAUSE WAS DETERMINED TO BE MANUFACTURING RELATED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
IT WAS REPORTED THAT SHORTLY BEFORE THE EXAMINATION, THE CATHETER WAS FOUND TO BE CONTAMINATED AS IT HAD UNIDENTIFIABLE PARTICLES FLOATING IN THE LIQUID. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S PHOTOGRAPHIC EVALUATION OF THREE IMAGES PROVIDED BY THE COSTUMER FOUND BLACK PARTICULATE MATTER SUSPENDED IN THE SALINE SOLUTION. BIOLOGICAL CONTAMINATION HAS BEEN IDENTIFIED IN PRIOR SYRINGES CONTAINING PARTICULATE CONTAMINATION.