XLUNG KIT 230
Received Jun 12, 2026 · Event occurred May 15, 2026
Report 3012172416-2026-00048 · MDR key 25484357
Device
Generic name
Extracorporeal System For Long-term Respiratory / Cardiopulmonary Failure
Manufacturer
Xenios AgCatalog number
32000014
Lot number
GSXH0416
Product problems
- Leak/Splash
- Mechanical Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
A USER FACILITY REPORTED THAT DURING THE ASSEMBLY AND PRIMING OF THE SUPPORT DEVICE, THE PRIMING SOLUTION LEAKED THROUGH THE PORTS FOR THE THERMAL UNIT. THE OXYGENATOR MEMBRANE WAS DAMAGED, SO THE CIRCUIT WAS THEREFORE UNUSABLE. THERE WAS NO PATIENT INVOLVEMENT. THE COMPLAINT SAMPLE WAS NOT AVAILABLE FOR PRODUCT EVALUATION.
Additional Manufacturer Narrative
PLANT INVESTIGATION: IT WAS REPORTED THAT DURING PRIMING OF THE XLUNG KIT 230, PRIMING SOLUTION LEAKED OUT OF THE PORTS OF THE THERMAL UNIT. AS THE OXYGENATOR MEMBRANE WAS FOUND DAMAGED AND UNUSABLE, THE KIT WAS REPLACED. A REVIEW OF THE BATCH DOCUMENTATION REVEALED NO NON-CONFORMITIES DURING THE MANUFACTURING PROCESS. NO OTHER SIMILAR COMPLAINT HAS BEEN REPORTED ABOUT THIS BATCH NUMBER. A DEFINITE CAUSE OF THE LEAK COULD NOT BE DETERMINED. BASED ON THE DESCRIPTION, POSSIBLE CAUSES COULD BE A DEFECT IN THE HEAT EXCHANGER MAT OR A DEFECT IN POTTING OF THE HEAT EXCHANGE CYLINDER. ALTHOUGH OUR OXYGENATORS ARE 100% LEAK TESTED, IN A FEW CASES LEAKAGE MAY OCCUR DUE TO ADVERSE ENVIRONMENTAL CONDITIONS OR MECHANICAL STRESS.