inforMED
MalfunctionLDF

TRANSLUMINAL BIPOLAR PACING PROBE

Received Jun 11, 2026 · Event occurred May 26, 2026

Report 2015691-2026-15983 · MDR key 25480417

Device

Generic name

Electrode, Pacemaker, Temporary

Model number

PE075F5

Lot number

NI

Product problems

  • Pacing Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE EVALUATION IS ANTICIPATED. A SUPPLEMENTAL REPORT WILL BE FORTHCOMING WHEN THE INVESTIGATION IS COMPLETED. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS A PART OF THE MONTHLY REVIEW.

Description of Event or Problem

IT WAS REPORTED THAT DURING USE OF THE SWAN-GANZ PACING CATHETER IT PROVIDED NOISE, A ZIGZAG PACING WAVEFORM, AND DID NOT PACE. THE ISSUE WAS RESOLVED BY REPLACING THE CATHETER. THERE WAS NO ALLEGATION OF PATIENT INJURY.