inforMED
MalfunctionLGD

ELECSYS TOXO IGG

Received Jun 11, 2026 · Event occurred May 18, 2026

Report 1823260-2026-02315 · MDR key 25476355

Device

Generic name

Enzyme Linked Immunoabsorbent Assay, Toxoplasma Gondii

Manufacturer

Roche Diagnostics

Catalog number

07028008190

Lot number

ASKU

Product problems

  • Non Reproducible Results

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE COBAS E 402 ANALYTICAL UNIT SERIAL NUMBER IS (B)(6) THE INVESTIGATION IS ONGOING.

Description of Event or Problem

ROCHE DIAGNOSTICS RECEIVED A CUSTOMER COMPLAINT REGARDING AN ALLEGED DISCREPANT ELECSYS TOXO IGG ASSAY RESULT FOR A PATIENT SAMPLE TESTED ON A COBAS PURE E 402 ANALYTICAL UNIT. THE SAME PATIENT'S SAMPLE ALIQUOT YIELDED 0.15 IU/ML ON A COBAS E 411, AND 4.30 IU/ML ON THE COMPETITOR DEVICE, WHILE IT YIELDED A NEGATIVE RESULT ON THE COBAS PURE E 402. A REACTIVE DISCREPANCY WAS ALSO REPORTED FOR THE SAME SAMPLE FOR ELECSYS TOXO IGM, BUT THE EXACT RESULTS WERE NOT PROVIDED.