SAFECLINITUBES
Received Apr 20, 2012 · Event occurred Feb 28, 2012
Report 3002807968-2012-00006 · MDR key 2546839
Device
Generic name
Plastic Capillary Tubes
Manufacturer
Radiometer Medical Aps.Model number
D941P-NA-240-85 X 3Catalog number
942-891
Lot number
DW 01
Product problems
- High Test Results
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
AN ABL835 ANALYZER REPORTED RESULTS OF BLOOD SAMPLE MEASUREMENTS OF NEONATAL PTS WITH AN INCREASED CA VALUE OF APP. 0, 3 MMOL/L HIGHER THAN EXPECTED WHEN MEASUREMENTS WERE PERFORMED ON CAPILLARY BLOOD SAMPLES USING RADIOMETER PLASTIC CAPILLARY TUBES.
Additional Manufacturer Narrative
ON THIS NEONATAL DEPARTMENT THEY HAVE HAD TWO DIFFERENT TYPES OF PLASTIC CAPILLARY TUBES, I.E. SAFECLINITUBES D941P-NA-240-85 X 3, PART NO. 942-891, AND SAFECLINITUBES D957P-70-100 X 1, PART NO. 942-892. THESE TWO TYPES OF PLASTIC CAPILLARIES WERE MIXED UP BY THE USERS (NURSES AND PHYSICIANS) WHEN BLOOD SAMPLE MEASUREMENTS HAVE BEEN PERFORMED. THEY HAVE USED THE CAPILLARY TYPE WHICH CONTAINS SODIUM HEPARIN WHEN THEY PERFORMED BLOOD SAMPLE MEASUREMENTS TO GET ELECTROLYTE RESULTS. THIS IS A USE ERROR. WE HAVE REVIEWED THE PACKAGE LABELING AND INSERT FOR THE SAFECLINITUBES (ATTACHED). BOTH THE PACKAGE AND THE INSERT STATES WHICH TYPE OF HEPARIN THE TUBES CONTAIN. FURTHERMORE THE ABL835 ANALYZER OPERATORS MANUAL CONTAINS THE FOLLOWING CAUTION (P. 1-5 'LIMITATIONS OF USE AND KNOWN INTERFERING SUBSTANCES,'): 'ANTICOAGULANTS THAT CONTAIN SODIUM SALTS WILL GIVE ERRONEOUSLY HIGH CNA+ RESULTS. SODIUM FLUORIDE WITH OR WITHOUT EDTA AND OXALATE (DI NA) INFLUENCE CGLU RESULTS. SODIUM FLUORIDE GIVES ERRONEOUSLY HIGH CNA+ AND LOW CCA2+, CGLU AND CLAC RESULTS. TRI SODIUM CITRATE INFLUENCES CNA+, CK+ AND CGLU RESULTS.'