inforMED
MalfunctionJCW

AMS 700 CX MS Pump

Received Jun 10, 2026 · Event occurred Mar 1, 2026

Report 2124215-2026-31059 · MDR key 25461075

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404233-12

Catalog number

72404233-12

Lot number

1100141344

Product problems

  • Degraded
  • Mechanical Problem
  • Material Puncture/Hole

Patient

64 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MODEL NUMBER/CATALOG NUMBER: 720185-01. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100146784. MODEL/CATALOG DESCRIPTION: RESERVOIR FLAT IZ 100 ML. UNIQUE IDENTIFIER (UDI) #: (B)(4). THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. BASED ON THE INFORMATION AVAILABLE THE CAUSE THAT CONTRIBUTED TO THE REPORTED EVENT CANNOT BE ESTABLISHED. THEREFORE, A CONCLUSION CODE OF CAUSE NOT ESTABLISHED WAS ASSIGNED TO THIS INVESTIGATION.

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT FUNCTIONING PROPERLY. THE PATIENT OBSERVED A BULGE AT THE BASE OF THE CYLINDER, WHICH SUBSEQUENTLY RUPTURED AND SNAPPED. FOLLOWING THIS EVENT, THE DEVICE BECAME NONFUNCTIONAL. THE DEVICE REMAINS IMPLANTED, AND A REPLACEMENT SURGERY HAS BEEN SCHEDULED. NO ADDITIONAL INFORMATION WAS PROVIDED.