TOTALCARE BARIATRIC PLUS BED
Received Jun 9, 2026 · Event occurred May 18, 2026
Report 1824206-2026-01314 · MDR key 25449221
Device
Generic name
Bed, Flotation Therapy, Powered
Manufacturer
Baxter Healthcare CorporationModel number
P1840DCatalog number
P1840DCA000010
Lot number
N/A
Product problems
- Activation Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE BAXTER TECHNICAL SUPPORT REPRESENTATIVE PERFORMED TROUBLESHOOTING OVER THE PHONE WITH THE ACCOUNT, ASKING THE ACCOUNT TO REPLACE THE CPR VALVE. PER THE BAXTER SERVICE MANUAL THE TOTALCARE BARIATRIC BED AND TOTALCARE BED SYSTEM REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM A SEMI-ANNUAL PREVENTATIVE MAINTENANCE. TEST THE CPR RELEASE FOR CORRECT OPERATION AND RESET OF THE HEAD CYLINDER. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. BAXTER TECHNICAL SUPPORT CONTACTED THE ACCOUNT ADDITION TIMES TRYING TO OBTAIN THE RESOLUTION FOR THIS EVENT. NO RESPONSE HAS BEEN RECEIVED FROM THE ACCOUNT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
BAXTER RECEIVED A REPORT THAT TOTALCARE BARIATRIC CAPITAL HAD CPR FUNCTION NOT WORKING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.