COBAS TAQMAN HCV 48 V2.0 CE-IVD QUANT

Received Apr 20, 2012 · Event occurred Mar 21, 2012

Report 2243471-2012-00016 · MDR key 2543770

Device

Generic name

Assay, Hybridization, Nucleic Acid Amplification For Detecti

Catalog number

04861817190

Lot number

058118

Product problems

  • High Test Results

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

A CUSTOMER FROM GREECE GENERATED DISCREPANT RESULTS WITH AN ACROMETRIX EXTERNAL CONTROL WHEN USING THE COBAS TAQMAN (CTM) (B)(6) TEST V.2.0 FOR USE WITH THE HIGH PURE SYSTEM COMPARED TO THE IN-HOUSE LIGHTCYCLER (LC) TEST. THE CTM (B)(6) V2.0 TEST GENERATED "TARGET NOT DETECTED" RESULTS. ONE OF THE SAMPLES, EXPECTED TO HAVE A TITER OF (B)(6), WAS TESTED WITH THE LC INHOUSE TEST AND GENERATED A RESULT OF (B)(6). THE OTHER TWO CONTROL SAMPLES GENERATED RESULTS OF (B)(6) WITH THE LC IN-HOUSE TEST. THE CTM (B)(6) V2.0 ROCHE KIT CONTROLS WERE VALID. IT IS NOT YET KNOWN, IF PATIENT RESULTS WERE INVOLVED.

Additional Manufacturer Narrative

THE INVESTIGATION INTO THIS REPORTED ISSUE IS ONGOING, AND THEREFORE, NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE CONCLUSION OF THIS INVESTIGATION WILL BE REPORTED THROUGH A FOLLOW-UP REPORT. (B)(4).

Additional Manufacturer Narrative

DEVICE EVALUATED BY MANUFACTURER: YES. RESULTS: DEVICE PERFORMED ACCORDING TO SPECIFICATIONS. CONCLUSION: NO FAILURE DETECTED AND PRODUCT PERFORMED WITHIN SPECIFICATION. NO PRODUCT OR BATCH NON-CONFORMANCE WAS IDENTIFIED. UPON INVESTIGATION, THERE WAS NO TREND FOUND IN THE FIELD. QC RELEASE DATA FOR 058118 WAS REVIEWED AND THE ISSUE OF DISCREPANT RESULTS HAS NOT BEEN OBSERVED. THERE HAVE NOT BEEN ANY MANUFACTURING NON-CONFORMANCE REPORTS FOR BATCH 058118. RETAIN TESTING OF THE COMPLAINT BATCH 058118 WAS PERFORMED ACCORDING TO KIT RELEASE PROCEDURES. ALL RESULTS WERE WITHIN SPECIFICATIONS; TESTING DID NOT REPRODUCE THE ALLEGATION. THE CUSTOMER REQUESTED SEQUENCING FOR THE AVAILABLE EXTERNAL CONTROLS. SAMPLES 128620 AND 128623 WERE SEQUENCED AND GENERATED THE FOLLOWING RESULTS: THERE WERE NO MISMATCHES PRESENT AT ANY OF THE (B)(4) FOR SAMPLES 128620 AND 128623, WHICH COULD IMPACT RESULTS. - SAMPLES 128620 AND 128623 WERE DETERMINED TO BE (B)(6). NO OTHER INFORMATION OR DATA REGARDING THE EXTERNAL CONTROLS WAS MADE AVAILABLE. (B)(4).