InjuryEWO

HEADBAND HB-7 FOR TDH39

Received Jun 8, 2026 · Event occurred Jun 4, 2026

Report 9612197-2026-00014 · MDR key 25424988

Add to watch list
Product code (EWO)
Model (1-25-12200)

Device

Generic name

Headband Hb-7 For Tdh39

Model number

1-25-12200

Catalog number

1-25-12200

Product problems

  • Break

Patient

NA · Unknown

  • Abrasion

Narrative

Additional Manufacturer Narrative

INITIAL REPORT FOR INTERNAL CASE NUMBER (B)(4). WE HAVE REQUESTED THE DEVICE TO BE RETURNED. RISK REVIEW- AS PER (B)(4) REV 15 - 1081 AURICAL AUD & MADSEN A450 - RISK ANALYSIS: HAZARD ID 6.5 CAUSE- MECHANICAL BREAKAGE, CAUSING SHARP CORNER, EDGES OR PINCH POINTS. E.G. DUE TO STEADY FORCE APPLIED ON THE DEVICE. (HIT BY SOMETHING). EFFECTS/HARM - MINOR PHYSICAL TRAUMA INJURIES. RESIDUAL RISK- 3 (LOW) AND ACCEPTABLE.

Description of Event or Problem

THE MANUFACTURER RECEIVED THE INFORMATION REGARDING THE HEADBAND HB-7 FOR TDH39 . THE USER HAS INFORMED ABOUT THE LOOSE HB-7 BRACKET. THE HB-7 BRACKET HAS COME LOOSE AND HIT PATIENT ON THE HEAD, RESULTING IN MINOR INJURIES. NO MEDICAL INTERVENTION WAS REQUIRED AT THIS TIME.

Remedial action

  • Replace