EXTENSION SET
Received Jun 5, 2026 · Event occurred May 6, 2026
Report 1416980-2026-01934 · MDR key 25417142
Device
Generic name
Filter, Infusion Line
Manufacturer
Baxter Healthcare CorporationModel number
2N3382Catalog number
2N3382
Lot number
NI
Product problems
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WAS ADDED TO D9, H3, H6, AND H11. H11: THE DEVICE AND A VIDEO OF THE SAMPLE WERE RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED TO THE ACTUAL SAMPLE RECEIVED, AND ALL COMPONENTS WERE CORRECTLY PLACED AND ACCORDING TO THE SPECIFICATION. NO VISUAL DEFECTS WERE OBSERVED IN THIS UNIT. THE SAMPLE WAS THEN SUBMITTED TO PRESSURE TESTING, AND A LEAK WAS NOT OBSERVED. IN ADDITION, THE SAMPLE WAS SUBMITTED TO CLEAR PASSAGE TESTING, AND NO BLOCKAGE WAS DETECTED. THE RETURNED VIDEO WAS REVIEWED, AND IT WAS NOTED THAT THERE WAS A LEAK FROM THE FILTER VENT HOLE. THE REPORTED CONDITION WAS VERIFIED IN THE VIDEO SAMPLE. THE CAUSE OF THE REPORTED CONDITION COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT A UNSPECIFIED QUANTITY OF MICRO-VOLUME EXTENSION SETS WITH 0.2 M FILTER WERE LEAKING FROM THE AIR VENTING MECHANISM. THE PROCESS STEP AT WHICH THE ISSUE OCCURRED WAS NOT SPECIFIED. IT WAS STATED THAT USERS MANUALLY PRIME THE LINES, OCCASIONALLY USING SYRINGES SMALLER THAN 10ML. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THESE EVENTS. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Additional Manufacturer Narrative
D4 UDI: AS THE LOT # IS UNKNOWN, THE UDI WILL CONSIST OF THE PRIMARY DI NUMBER ONLY. (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.