inforMED
MalfunctionDXO

TRUWAVE DISPOSABLE PRESSURE TRANSDUCER

Received Jun 5, 2026

Report 2015691-2026-15820 · MDR key 25404662

Device

Generic name

Transducer, Pressure, Catheter Tip

Model number

PX2X2

Lot number

66643543

Product problems

  • Incorrect Measurement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

AS REPORTED, THIS PRESSURE MONITORING SET HAD ZEROING DIFFICULTIES. ONCE THE ZERO COULD BE COMPLETED, THE ARTERIAL PRESSURE VALUES PROVIDED WERE NOT CORRECT (MANUFACTURER REFERENCE (B)(4) / FDA MDR REFERENCE 2015691-2026-15819). THEN, THE DEVICE WAS REPLACED; HOWEVER, SAME ISSUE OCCURRED WITH THE SECOND UNIT (MANUFACTURER REFERENCE (B)(4)). FINALLY, ISSUE WAS SOLVED USING A THIRD DEVICE. NO ERROR MESSAGE NOR ABNORMAL WAVEFORM WERE OBSERVED. THERE WERE NO FURTHER DETAILS AVAILABLE. THERE WAS NO ALLEGATION OF PATIENT INJURY. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

Additional Manufacturer Narrative

THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS; HOWEVER, IT HAS NOT YET BEEN RECEIVED. UPON THE RETURN OF THE PRODUCT A SUPPLEMENTAL REPORT WILL BE SENT WITH THE INVESTIGATION RESULTS. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW.