inforMED
MalfunctionJCF

BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE

Received Jun 4, 2026 · Event occurred Apr 23, 2026

Report 1917413-2026-00498 · MDR key 25403193

Device

Generic name

Lymphocyte Separation Medium

Catalog number

362761

Lot number

5230368

Product problems

  • Short Fill

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: BD RECEIVED THREE PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR UNDERFILL WAS OBSERVED. ADDITIONALLY, 60 RETAINED SAMPLES WERE VISUALLY INSPECTED, AND THE HEMOGARD CLOSURE ASSEMBLY WAS FOUND TO BE CORRECTLY ASSEMBLED AND PLACED ON THE TUBES. FUNCTIONAL TESTS WERE CONDUCTED ON 10 RETAINED SAMPLES, AND ALL TUBES WERE WITHIN SPECIFICATION LIMITS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE UNDERFILL BASED ON CUSTOMER PHOTOS. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem

IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE TUBES, THREE (3) TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.