MalfunctionDQY
EXCELSIOR XT-17 STD STRAIGHT 150CM
Received Jun 4, 2026 · Event occurred Mar 19, 2026
Report 3008881809-2026-00349 · MDR key 25398247
Device
Generic name
Catheter, Percutaneous
Manufacturer
Stryker Neurovascular CorkCatalog number
M003C1775ST0
Lot number
25865985
Product problems
- Material Puncture/Hole
Patient
57 YR · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE MICROCATHETER (SUBJECT DEVICE) WAS RETURNED WITH THE ATLAS STENT DEVICE AND THE VISUAL/MICROSCOPIC INSPECTION REVEALED THAT THE SDW (STENT DELIVERY WIRE) OF THE ATLAS STENT HAD PIERCED THROUGH THE LUMEN/SHAFT OF THE SUBJECT MICROCATHETER. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Additional Manufacturer Narrative
THIS IS 2 OF 2 MDRS.