inforMED
MalfunctionEXE

NATURA

Received Jun 4, 2026

Report 9618003-2026-01050 · MDR key 25389237

Device

Generic name

Protector, Ostomy

Model number

411803

Lot number

5H04955

Product problems

  • Malposition of Device

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEVICE 10 OF 10. BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A REPORTABLE MALFUNCTION. BATCH RECORD REVIEW: LOT 5H04955 WAS MANUFACTURED ON 29/AUG/2025, IN ATS LINE, WITH A TOTAL OF (B)(4) MARKET UNITS (MKUS). THE COMPLAINTS ENGINEER I PERFORMED A BATCH RECORD REVIEW ON 225/MAY/2026, TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED AND NO ISSUES WERE FOUND; ALL THE COMPONENTS FOR ASSEMBLY WERE CORRECT PER BILL OF MATERIALS (BOM) AND ALL THE TOOLING INFORMATION DOCUMENTED WAS ALSO CORRECT, SYSTEM APPLICATION PRODUCT (SAP) MATERIAL IDENTIFICATION (ID) 1222274 AND MANUFACTURING ORDER (B)(4). REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WAS FOUND WITHIN THE DOCUMENTATION. THE DEFECT REPORTED BY THE CUSTOMER COULD OCCUR DURING THE SUB-ASSEMBLY PROCESS PERFORMED IN THE ATS 2 LINE. FOR THIS REASON, A DETAILED BATCH RECORDS REVIEW WAS PERFORMED OF THE SUBASSEMBLY LOT MANUFACTURED IN THIS LINE. THE CENTER HOLE CUTTING PROCESS IS PERFORMED IN THE DELTA CRUSADER: THE SUBASSEMBLY LOT: 5H04864, ORDER (B)(4), MATERIAL 1220484, WAS MANUFACTURED ON 09/01/2025 IN THE DELTA CRUSADER NEW MANUFACTURING LINE, WITH A TOTAL OF (B)(4) EACH (EA). THE COMPLAINT ENGINEER PERFORMED A BATCH RECORD REVIEW ON 05/25/2026, AND IT WAS CONFIRMED THAT THE APPLICABLE PROCEDURES WERE FOLLOWED, THE MATERIALS WERE CORRECT AS PER BOM, AND THE EQUIPMENT SETTINGS WERE ACCORDING TO THE PROCESS INSTRUCTION. THE SUBASSEMBLY LOT: 5J01942, ORDER (B)(4), MATERIAL 1220484, WAS MANUFACTURED ON 09/16/2025 IN THE DELTA CRUSADER NEW MANUFACTURING LINE, WITH A TOTAL OF (B)(4) EA. THE COMPLAINT ENGINEER PERFORMED A BATCH RECORD REVIEW ON 05/25/2026, AND IT WAS CONFIRMED THAT THE APPLICABLE PROCEDURES WERE FOLLOWED, THE MATERIALS WERE CORRECT AS PER BOM, AND THE EQUIPMENT SETTINGS WERE ACCORDING TO THE PROCESS INSTRUCTION. THE SUBASSEMBLY LOT: 5H04862, ORDER (B)(4), MATERIAL 1220484, WAS MANUFACTURED ON 09/01/2025 IN THE DELTA CRUSADER NEW MANUFACTURING LINE, WITH A TOTAL OF (B)(4) EA. THE COMPLAINT ENGINEER PERFORMED A BATCH RECORD REVIEW ON 05/25/2026, AND IT WAS CONFIRMED THAT THE APPLICABLE PROCEDURES WERE FOLLOWED, THE MATERIALS WERE CORRECT AS PER BOM, AND THE EQUIPMENT SETTINGS WERE ACCORDING TO THE PROCESS INSTRUCTION. PHOTOGRAPH, VIDEO AND/OR PHYSICAL SAMPLE EVALUATION: THERE WERE PHOTOGRAPHS ASSOCIATED WITH THIS CASE AND IN THESE, THE REPORTED DEFECT CAN BE SEEN. NO UNUSED RETURN SAMPLE WAS EXPECTED. HISTORICAL COMPLAINTS REVIEW: ON 25/MAY/2026, COMPLAINTS ENGINEER RAN A QUERY IN DATABASE IN ORDER TO VERIFY THE COMPLAINTS REPORTED FOR THE 5H04955 LOT FOR THE MALFUNCTION ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (MOLDABLE ONLY)¿ DEFECT AND NO ADDITIONAL COMPLAINTS WERE IDENTIFIED. HISTORICAL NONCONFORMANCE REVIEW: ON 5/MAY/2026, COMPLAINTS ENGINEER RAN A QUERY IN DATABASE LOOKING FOR ANY IN PROCESS NONCONFORMANCE / CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) (S) ASSOCIATED TO THE MALFUNCTION ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (MOLDABLE ONLY)¿ DEFECT FOR THE LOT NUMBER 5H04955 AND AS RESULT, NO NONCONFORMANCE / CAPA (S) FOR THIS MALFUNCTION WERE GENERATED DURING THE MANUFACTURING PROCESS OF THE REFERENCED LOT. CURRENT QUALITY CONTROLS: BASED ON THE PROCESS INSTRUCTION (PI), THE FOLLOWING TESTS ARE PERFORMED IN THE MANUFACTURING LINES, IN ORDER TO IDENTIFY THIS FAILURE MODE IN OUR MANUFACTURING PROCESS: TEST METHODS (TM) ¿PACKAGE SEAL INTEGRITY NONCONFORMITIES - METHOD 8¿ (B)(4). DIMENSIONAL INSPECTION: - FILM RELEASE LINER: THE FILM MUST REMAIN INTACT ALONG THE LENGTH OF THE WAFER WITHOUT ANY CUTS IN THE CENTER HOLE. THE FILM RELEASE TAB MUST BE FREE AND NOT BE TRAPPED UNDER THE ADHESIVE DISK. (B)(4). - COLLAR CONCENTRICITY: SUPPORT RING CONCENTRIC WITHIN 2.0 MM TO THE COLLAR THROUGH THE CUT. CONCENTRICITY METER: (45MM FLANGE). (B)(4). - ADHESIVE DISC ALIGNMENT: THE DISC MUST NOT CONTACT THE SRP ADHESIVE DISC ID COLLAR AND MUST BE CONCENTRIC WITHIN 2.0MM. (B)(4). DEFECT RATE ANALYSIS: THERE HAVE BEEN ONLY 10 DEFECTIVE PARTS CONFIRMED TO DATE FROM A LOT SIZE OF (B)(4) PRODUCTS. THIS REPRESENTS A DEFECT RATE OF (B)(4) PERCENT WHICH WAS WELL WITHIN AN APPROPRIATE ACCEPTABLE QUALITY LEVEL (AQL) FOR LEAKAGE WHICH SHOULD BE (B)(4) PERCENT BASED ON OUR STANDARD OPERATING PROCEDURE (SOP). IN ADDITION, ALL THE IN-PROCESS TESTING ON THIS LOT DID NOT FIND A SINGLE DEFECTIVE UNIT, WHICH CONFIRMS THAT THE LOT WAS UNLIKELY TO BREACH AN AQL OF (B)(4). IMPORTANTLY TO DATE, IT WAS WELL WITHIN OUR ACCEPTED AQL LEVEL FOR THIS TYPE OF FAILURE MODE OR DEFECT. CONCLUSIONS: A REVIEW OF BATCH RECORDS WAS COMPLETED AND SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCIES RELATED TO THIS ISSUE WERE FOUND WITHIN THE DOCUMENTATION. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET PRODUCT MONITORING REVIEW PROCESS, STANDARD OPERATING PROCEDURE (SOP). TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER: REPORTING SITE: 1049092. MANUFACTURING SITE: 9618003.

Description of Event or Problem

THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.