TOTALCARE BARIATRIC PLUS BED
Received Jun 3, 2026 · Event occurred May 11, 2026
Report 1824206-2026-01278 · MDR key 25385850
Device
Generic name
Bed, Flotation Therapy, Powered
Manufacturer
Baxter Healthcare CorporationModel number
P1840DCatalog number
P1840DCA000010
Product problems
- Positioning Failure
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE HEAD UP WAS NOT WORKING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Additional Manufacturer Narrative
THE BAXTER TECHNICIAN FOUND THE SOLENOID VALVE NEEDED TO BE REPLACED. PER THE BAXTER SERVICE MANUAL THE TOTALCARE BED SYSTEM REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM A SEMI-ANNUAL PREVENTATIVE MAINTENANCE. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. TEST THE HEAD SECTION SLIDE PIVOTS, SELF-LUBRICATING BEARING, AND SLIDER PIVOT EXTRUSIONS. LOOK FOR DAMAGE, AND MAKE SURE THE HEAD SECTION OPERATES CORRECTLY. REPLACE COMPONENTS AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN MAY 06, 2026. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE SOLENOID VALVE TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.