AMS 700 CX MS PUMP
Received Jun 2, 2026 · Event occurred May 6, 2026
Report 2124215-2026-29576 · MDR key 25368537
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
72404232-10Catalog number
72404232-10
Lot number
1000532032
Product problems
- Fluid/Blood Leak
- Inflation Problem
Patient
85 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. FLUID LEAK AND INFLATION ISSUE ARE ACKNOWLEDGED WITHIN THE INSTRUCTIONS FOR USE (IFU) AND ARE NOT RELATED TO OFF-LABEL USE OR FAILURE TO FOLLOW INSTRUCTIONS. BASED ON THE INFORMATION AVAILABLE THE CAUSE THAT CONTRIBUTED TO THE REPORTED EVENT CANNOT BE ESTABLISHED; THEREFORE, A CONCLUSION CODE OF CAUSE NOT ESTABLISHED WAS ASSIGNED TO THIS INVESTIGATION.
Description of Event or Problem
IT WAS REPORTED THAT A PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED WITH A NON INFLATING DEVICE. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PHYSICIAN IDENTIFIED FLUID LOSS FROM THE LEFT CYLINDER. THE ORIGINAL RESERVOIR WAS DRAINED AND RETAINED, AND ALL OTHER COMPONENTS WERE EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS.