SURPASS ELITE FDS 4.0MM X 17MM - PMAS
Received Jun 2, 2026 · Event occurred May 19, 2026
Report 3008881809-2026-00345 · MDR key 25364987
Device
Generic name
Intracranial Aneurysm Flow Diverter
Manufacturer
Stryker Neurovascular CorkCatalog number
FDSE40017
Lot number
25206433
Product problems
- Premature Activation
- Difficult to Remove
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE STENT (SUBJECT DEVICE) WAS USED IN THE MEDICAL PROCEDURE. HOWEVER, THE SUBJECT STENT PREMATURELY DETACHED FROM THE RESHEATH PAD WHILE RECAPTURING DEVICE. THE PHYSICIAN REPORTED RESISTANCE WHILE RESHEATHING AND FELT A POP. THE DELIVERY WIRE WAS MOVING INDEPENDENTLY OF THE STENT INDICATING PREMATURE STENT DEPLOYMENT. THE PHYSICIAN PULLED THE MICROCATHETER (WITH HALF OF THE STENT WITHIN THE MICROCATHETER AND ANOTHER HALF OUTSIDE) INTO THE GUIDE CATHETER AND REMOVED THEM. THE STENT WAS FOUND DEPLOYED IN THE HUB OF THE GUIDE CATHETER. THE PROCEDURE WAS DELAYED BY 60 MINUTES. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS REPORTED TO BE COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT,