inforMED
MalfunctionOUT

SURPASS ELITE FDS 4.0MM X 17MM - PMAS

Received Jun 2, 2026 · Event occurred May 19, 2026

Report 3008881809-2026-00345 · MDR key 25364987

Device

Generic name

Intracranial Aneurysm Flow Diverter

Catalog number

FDSE40017

Lot number

25206433

Product problems

  • Premature Activation
  • Difficult to Remove

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE STENT (SUBJECT DEVICE) WAS USED IN THE MEDICAL PROCEDURE. HOWEVER, THE SUBJECT STENT PREMATURELY DETACHED FROM THE RESHEATH PAD WHILE RECAPTURING DEVICE. THE PHYSICIAN REPORTED RESISTANCE WHILE RESHEATHING AND FELT A POP. THE DELIVERY WIRE WAS MOVING INDEPENDENTLY OF THE STENT INDICATING PREMATURE STENT DEPLOYMENT. THE PHYSICIAN PULLED THE MICROCATHETER (WITH HALF OF THE STENT WITHIN THE MICROCATHETER AND ANOTHER HALF OUTSIDE) INTO THE GUIDE CATHETER AND REMOVED THEM. THE STENT WAS FOUND DEPLOYED IN THE HUB OF THE GUIDE CATHETER. THE PROCEDURE WAS DELAYED BY 60 MINUTES. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS REPORTED TO BE COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT,