inforMED
MalfunctionHTN

KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLAN

Received Jun 2, 2026 · Event occurred May 8, 2026

Report 1220246-2026-03114 · MDR key 25359381

Device

Generic name

Washer, Bolt, Nut

Manufacturer

Arthrex, Inc.

Catalog number

AR-2372BLO

Lot number

15542833

Product problems

  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Description of Event or Problem

ON 13-MAY-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-2372BLO KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLANT EXPERIENCED AN ISSUE IN WHICH THE BUTTON BROKE OFF THE INSERTER BUT REMAINED ATTACHED TO THE SUTURE. THIS WAS IDENTIFIED DURING AN AC JOINT RECONSTRUCTION PROCEDURE PERFORMED ON (B)(6) 2026. NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 21-MAY-2026: THE ISSUE OCCURRED DURING INSERTION INTO THE PATIENT. THE AFFECTED BUTTON AND SUTURES WERE REMOVED, AND ALL DETACHED FRAGMENTS WERE SUCCESSFULLY RETRIEVED. THE PROCEDURE WAS COMPLETED USING ANOTHER AR-2372BLO KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLANT. THERE WAS AN APPROXIMATE DELAY OF 5-10 MINUTES, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.