inforMED
MalfunctionCAL

PENTAX

Received Jun 2, 2026 · Event occurred May 21, 2026

Report 9610877-2026-51220 · MDR key 25356775

Device

Generic name

Fiberopticintubationscope 2.6c 5.1tp Rop

Model number

FI-16BS

Product problems

  • Loose or Intermittent Connection

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

G4: THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. D4 UDI: "NOT DISTRIBUTED IN USA", BECAUSE PRODUCTS ARE NOT DISTRIBUTED IN USA PRODUCTS, BUT SIMILAR DEVICE PRODUCT ARE LISTED IN USA. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE BIOPSY INLET PIECE LOOSE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE BIOPSY INLET PIECE. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE ETHYLENE OXIDE GAS VALVE (EOG) BROKEN, AND THE INSERTION FLEXIBLE TUBE BUCKLED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT "HR-RPT-0620(CONTROL BODY)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. BIOPSY INLET PIECE LOOSENED.