inforMED
MalfunctionHWQ

FIRSTPASS SUTURE PASSER

Received Jun 1, 2026 · Event occurred Apr 28, 2026

Report 3006524618-2026-00281 · MDR key 25355780

Device

Generic name

Passer

Catalog number

22-4035

Lot number

K5001AF

Product problems

  • Detachment of Device or Device Component

Patient

56 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H11: INTERNAL COMPLAINT REFERENCE: (B)(4). H3, H6 THIS COMPLAINT WAS OPENED BY SMITH+NEPHEW TO DOCUMENT A PRODUCT PROBLEM ASSOCIATED WITH A SMITH+NEPHEW DEVICE. THE REPORTED PROBLEM RELATES TO KNOWN RISKS WITH THE DEVICE ITSELF OR WITH ITS USE THAT ARE APPROPRIATELY DOCUMENTED IN OUR RISK FILES. SMITH+NEPHEW WILL CONTINUE TO MONITOR TRENDS IN ACCORDANCE WITH OUR POST-MARKET SURVEILLANCE PROCESS AND TAKE NECESSARY ACTION AS REQUIRED IF ANTICIPATED SEVERITY AND/OR OCCURRENCE RATES ARE EXCEEDED. SMITH+NEPHEW HAS NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY SPECIFICATIONS AT THE TIME OF MANUFACTURE. BASED ON OUR REVIEW OF ALL CURRENTLY AVAILABLE INFORMATION, WE ARE UNABLE TO IDENTIFY A DEFINITIVE ROOT CAUSE. HOWEVER, AS THE USE OF OUR PRODUCT CANNOT BE EXCLUDED AS A POTENTIAL CAUSE OR CONTRIBUTORY FACTOR TO THE REPORTED ISSUE, WE ARE SUBMITTING THIS REPORT IN ACCORDANCE WITH APPLICABLE REGULATIONS. IF ADDITIONAL INVESTIGATIVE FINDINGS OR INFORMATION BECOMES AVAILABLE THAT ALTERS THE CONCLUSIONS OF THIS REPORT, A FOLLOW-UP REPORT WILL BE SUBMITTED AS REQUIRED.

Description of Event or Problem

IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY ROTATOR CUFF, WHEN THE JAW OF THE FORCEPS WAS OPENED, THE JAWBONE DETACHED FROM THE FORCEPS AND WAS RECOVERED. NO PIECES FELL INTO THE PATIENT'S WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR SN BACK-UP. THE PATIENT WAS STABLE. NO FURTHER COMPLICATIONS WERE REPORTED.