MEDLINE
Received Jun 1, 2026 · Event occurred Apr 1, 2026
Report 1417592-2026-01949 · MDR key 25354665
Device
Generic name
1lyr Try 16fr 10ml 100% Sili Um Snap
Manufacturer
Medline Industries, LpCatalog number
URO175816
Product problems
- Deflation Problem
Patient
NA · Unknown
- Genital Bleeding
Narrative
Additional Manufacturer Narrative
IT WAS REPORTED THAT UPON DEFLATION OF A CATHETER, "THERE WAS STILL A TINY AMOUNT OF AIR INFLATED IN THE BALLOON". THE CUSTOMER CONTACT REPORTED, "CONNECTED A STERILE SYRINGE TO THE BALLOON PORT AND ALLOWED 9.5ML TO NATURALLY BACKFLOW INTO THE SYRINGE. THEN PULLED BACK WITH THE SYRINGE TO DOUBLE-CHECK IF THERE WAS ANY MORE AIR OR WATER LEFT IN THE BALLOON. PULLED BACK 2 ADDITIONAL TIMES TO VERIFY THERE WAS NOTHING LEFT IN THE BALLOON. CATHETER PULLED OUT WITH NO RESISTANCE. WHEN IT WAS COMPLETELY OUT, THERE WAS STILL A TINY AMOUNT OF AIR INFLATED IN THE BALLOON. PT BLED FROM HIS URETHRA FOR A FEW MINUTES AFTERWARDS AND THEN RESOLVED". THERE WERE NO REPORTS OF DEATH, SERIOUS INJURY, MEDICAL INTERVENTION, OR FOLLOW-UP CARE. IT HAS BEEN DETERMINED THAT THE EVENT DID NOT INVOLVE A DEATH OR SERIOUS INJURY; HOWEVER, IT MEETS THE DEFINITION OF A REPORTABLE MALFUNCTION, AS RECURRENCE OF THE MALFUNCTION COULD CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY. BASED ON THE INFORMATION REVIEWED, THIS IS A REPORTABLE EVENT AS DEFINED UNDER 21 CFR 803.3.
Description of Event or Problem
IT WAS REPORTED THAT UPON DEFLATION OF A CATHETER, "THERE WAS STILL A TINY AMOUNT OF AIR INFLATED IN THE BALLOON".