inforMED
MalfunctionCJY

ALINITY C CALCIUM2

Received Jun 1, 2026 · Event occurred Apr 1, 2026

Report 3005094123-2026-00258 · MDR key 25354418

Device

Generic name

Azo Dye, Calcium

Catalog number

04T87-30

Lot number

82017UD00

Product problems

  • Imprecise or Erratic Results

Patient

68 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.

Description of Event or Problem

THE CUSTOMER OBSERVED IMPRECISE ALINITY CALCIUM2 RESULTS FOR ONE 68-YEAR-OLD MALE PATIENT. PRESENTATION: ITU POST VASCULAR PROCEDURE FOR PERIPHERAL VASCULAR DISEASE AND DIABETIC FOOT INFECTION, T2DM, HAEMORRHAGE, ACUTE HYPERCAPNIC RESPIRATORY FAILURE, AKI, ALCOHOLIC CIRRHOSIS AND PORTAL HYPERTENSTION, ESSENTIAL HYPERTENSION, IRON DEFICIENCY ANAEMIA. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL RESULT = 3.05 REPEATS 3.67, 2.65, 2.37, 2.77, 2.29, 3.38, 2.62, 2.48, 2.31, 2.43, 3.02, 3.33, AND 3.25 MMOL/L. MEDICATIONS: HUMULIN I, NOVORAPID INJECTION, GLYCEROL, FUROSEMIDE, LACTULOSE, MACROGOL, MIRCERA INJECTION, CARVEDILOL, PREDNISOLONE, BUPRENORPHINE PATCH, OCTENIDINE, SENNOSIDES, ANUSOL HC OINTMENT, CETIRIZINE, CHOLECALCIFEROL [SIC], LANSOPRAZOLE, PHOSPHATE ENEMA, CHLORPHENAMINE INJECTION, OXYCODONE ORAL, ENOXAPARIN EVERY EVENING SC, PARACETAMOL, LINAGLIPTIN, NALOXONE, ONDANSETRON, CYCLIZINE, RIFAXIMIN. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.