STRYKER GROUNDING PAD
Received Jun 1, 2026
Report MW5188747 · MDR key 25348174
Device
Generic name
Generator, Lesion, Radiofrequency
Manufacturer
Stryker Leibinger Gmbh & Co. Kg.Catalog number
0406-650-210
Lot number
202709LX
Product problems
- Connection Problem
- Therapeutic or Diagnostic Output Failure
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
AT THE PAIN CLINIC, ON (B)(6) 2026, WE HAD A PROCEDURE. CONNECTED GROUNDING PAD (PRODUCT # 0406-650-210) TO CABLE, NOT REGISTERED. CHANGE PAD, SAME RESULT. CHANGED CABLE W/ SAME RESULT. TRIED DIFFERENT MACHINE W/SAME RESULT. AT THIS TIME, LOOKED FOR DIFFERENT PAD- HAD SOME OLD STOCK OF 0406-650-205. CONNECTED PAD; WORKED AS INTENDED. PULLED ALL STOCK WITH SAME LOT NUMBER 202709LX. PROCEED TO DO PROCEDURE WITH OLD PAD, WORKED EFFECTIVELY. NO HARM TO PATIENT DONE. IN THE EVENT WE DIDN'T HAVE SOME OF THE OLD STOCK, WE WOULD'VE HAD TO CANCEL 3-4 PATIENTS TODAY, AND UNTIL FURTHER NOTICE. ALL PHYSICIANS, CNOIC, NCOIC WERE NOTIFIED. (B)(4) WAS INITIATED BY (B)(6), CNOIC. I BELIEVE THIS ITEM IS DEFICIENT. THIS WAS DISCOVERED TODAY. OUR MAIN SUPPLIER I UNDERSTAND IS (B)(4), AND THERE ARE NO SUBSTITUTIONS AVAILABLE THROUGH GOVERNMENT CONTRACTED SUPPLIERS. STRYKER IS THE RF GENERATOR MANUFACTURER, AND SUPPLIES (B)(4) WITH THIS PRODUCT. THE PREVIOUS VERSION OF THIS SAME PRODUCT PROVED TO BE STABLE. HOWEVER, THIS NEW VERSION OF THEIR PRODUCT, IN ADDITION TO BEING ALMOST 4X MORE EXPENSIVE, SEEMS NOT AS EFFECTIVE AS THE PREVIOUS VERSION. AT LEAST 10 PC TESTED. 85 ON HAND TO TURN IN. WO# (B)(4) INITIATED WITH MED MAINT. PT: 4582. DEVICE: 3023, 2900. REFERENCE REPORT: MW5188744, MW5188745, MW5188746, MW5188748, MW5188749, MW5188750, MW5188751, MW5188752, MW5188753.