IMPAX CV DICOMSTORE WITH MEDIA PURGE DAEMON
Received Apr 5, 2011 · Event occurred Feb 13, 2012
Report 1225058-2012-00005 · MDR key 2534733
Device
Generic name
Medical Image Storage Device
Manufacturer
Agfa Healthcare CorpModel number
ALL VERSIONSCatalog number
NA
Lot number
NA
Product problems
- Computer Software Problem
- Inadequate Storage
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
(MEDIA PURGE DAEMON (MDP) IS AN OLDER AGFA PRODUCT USED TO PURGE THE IMAGES STORED ON THE PRIMARY MEMORY CACHE ONCE THEY HAVE BEEN ARCHIVED TOP A LONG TERM ARCHIVE AND THE AMOUNT OF DATA STORED ON-LINE REACHES A PREDEFINED AMOUNT. MPD ALLOWS THE SYSTEM TO CONTINUOUSLY RECEIVE THE RECENTLY ACQUIRED IMAGES FROM THIRD PARTIES MODALITIES.THIS TOOL WAS DISCONTINUED AND REPLACED BY CARDIO PURGE SERVICE (CPS) IN (B)(4) 2008.) THE CUSTOMER SITE WAS STILL USING MPD AT THE TIME OF THIS EVENT. MPD HAS SINCE BEEN DEACTIVATED AT THE SITE AND WILL BE REPLACED BY IMPAX CV CARDIO PURGE SERVICE. AGFA SERVICE HAS RECOVERED THE LOST IMAGES.
Description of Event or Problem
AGFA SUBMITTED MDR REPORT # 1225058-2012-00003 TO THE FDA ON (B)(4) 2012. ON (B)(6) 2012, A DIFFERENT CUSTOMER REPORTED THE SAME ISSUE OF NOT BEING ABLE TO ACCESS OLDER IMAGES FROM THEIR SAN (STORAGE AREA NETWORK). THE CUSTOMER STATED THAT ULTRASOUND IMAGES COULD NOT BE PUSHED FROM THE ULTRASOUND CARTS TO IMPAX CV, AS THE DISKS OF IMPAX CV WERE FULL AND IMAGES FROM YRS 2009 AND 2010 COULD NOT BE ACCESSED ANYMORE. AGFA DETERMINED WRONG CONFIGURATION OF DICOMSTORE IN COMBINATION WITH MEDIA PURGE DAEMON IS THE ROOT CAUSE OF THIS EVENT. THE CUSTOMER CONFIRMED THERE WERE NO REPORTS OF PT HARM, DUE TO THE UNAVAILABILITY OF THE ULTRASOUND IMAGES AS ALL REPORTS OF THE IMAGES WERE AVAILABLE.
Remedial action
- Repair