inforMED
InjuryOXC

NEOVEIL

Received May 31, 2026 · Event occurred Jan 30, 2026

Report 3012602431-2026-00046 · MDR key 25344872

Device

Generic name

Mesh, Surgical, Absorbable, Staple Line Reinforcement, Product Code: Oxc

Manufacturer

Gunze Limited

Model number

NV-ET-M60A-2

Catalog number

NV-ET-M60A-2

Product problems

  • Material Integrity Problem

Patient

NA · Unknown

  • Foreign Body In Patient
  • Insufficient Information

Narrative

Description of Event or Problem

A CUSTOMERS USES NEOVEIL TUBES IN BARIATRIC SURGERY. HE RECENTLY TESTED REFERENCE NV-ET-M60A-2 WITH GOLD PANTHER STAPLERS CADF-60S (GOLD). DURING STAPLING, A WIRE BROKE, AND ONE OF THE WHITE PARTS REMAINED SEWN TO THE BLUE PART. THE SURGEON CUT THE WHITE PART TO REMOVE IT, BUT A SMALL FRAGMENT WAS LEFT IN THE PATIENT. BASED ON OUR OBSERVATIONS, THIS ISSUE MAY BE DUE TO A CERTAIN FRAGILITY OF THE NV TUBE WIRES.