GENTUITY HF-OCT IMAGING SYSTEM

Received May 28, 2026 · Event occurred Dec 12, 2024

Report 3011578544-2026-00031 · MDR key 25320788

Device

Generic name

System, Imagin, Optical Coherence Tomography (oct)

Manufacturer

Gentuity, Llc

Model number

G10-01

Catalog number

G10-01

Product problems

  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

A RECENT CONVERSATION AMONGST CLINICAL SPECIALISTS PROMPTED ME TO REVIEW EACH RUN IN THE CONSOLE FOR REPETITIVE/STACKED FRAMES AND I FOUND THAT IN 26 OF THE 28 RUNS FROM 16 CASES HERE HAD DISTAL FRAMES STACKED/FAILURE TO PULL BACK BEFORE GENERATING IMAGES.

Additional Manufacturer Narrative

REVIEW OF 16 CASES FOUND DISTAL REPEATED FRAMES IN 26 OF 28 RUNS (¿8¿10 MM). AN ENGINEERING STUDY VALIDATED THIS REPEAT BUFFER AS AN INTENTIONAL DESIGN CHOICE TO ENSURE DISTAL COVERAGE. NO SOFTWARE, HARDWARE, OR CATHETER FAILURES WERE IDENTIFIED. THE SYSTEM IS FUNCTIONING AS DESIGNED.