ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2
Received May 28, 2026 · Event occurred May 4, 2026
Report 1823260-2026-02107 · MDR key 25319402
Device
Generic name
Nitrophenylphosphate, Alkaline Phosphatase Or Isoenzymes
Manufacturer
Roche DiagnosticsCatalog number
08056757190
Lot number
93559301
Product problems
- High Test Results
- Non Reproducible Results
Patient
76 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE ANALYZER'S SERIAL NUMBER IS (B)(6). THE FIELD SERVICE REPRESENTATIVE PERFORMED ADJUSTMENTS, PERFORMED CHECKS, CLEANED THE FLOW PATHS, AND PERFORMED CHECKS AND TESTS WITH ACCEPTABLE RESULTS. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 110 U/L, AND THE REPEATED RESULT WAS 74 U/L. THE SAMPLE WAS REPEATED ON ANOTHER MODULE, AND THE RESULT WAS 73 U/L. THE SAMPLE WAS REPEATED BECAUSE THE RESULT DIDN'T MATCH THE PATIENT'S CLINICAL PICTURE. THE REPEATED RESULT (73 U/L) WAS BELIEVED TO BE CORRECT.