inforMED
InjuryBXB

HOCOMA

Received May 28, 2026 · Event occurred May 22, 2026

Report 3003781275-2026-00001 · MDR key 25318307

Device

Generic name

Lokomatpro L6.2 Freed Kombi

Manufacturer

Hocoma Ag

Catalog number

30682

Product problems

  • Use of Device Problem

Patient

73 YR · Female

  • Bone Fracture(s)

Narrative

Additional Manufacturer Narrative

LOOSING A SHOE IS HARDLY POSSIBLE WHEN USER STICKS TO THE BEST PRACTICE FOR THIS DEVICE. IT LEADS US TO THE CONCLUSION THAT THE FOOTLIFTERS OF THE DEVICE WERE NOT USED DESPITE BEING RECOMMENDED BEST PRACTICE BY THE MANUFACTURER.

Description of Event or Problem

PATIENT LOOSES SHOE AND DRAGS TOES ON TREADMILL OF ROBOTIC REHABILITATION DEVICE. LOKOMAT WAS STOPPED BY THERAPIST AND TRAINING CONTINUED. X-RAY ON (B)(6) 26. REVEALED DISTAL FRACTURES OF TIBIA AND FIBULA.