VIDEO-OPTIK "ENDOEYE 3D", 30°
Received May 28, 2026 · Event occurred Apr 20, 2026
Report 9610773-2026-04115 · MDR key 25316660
Device
Product problems
- Poor Quality Image
- Failure to Clean Adequately
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: DUE TO THE COVER GLASS OF THE INSERTION SECTION CONTAINED FOREIGN MATERIAL, THE IMAGE QUALITY WAS POOR. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem
THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION. DURING TESTING, THE SERVICE REPAIR TECHNICIAN IDENTIFIED THE DEVICE'S COVER GLASS OF THE INSERTION SECTION CONTAINED FOREIGN MATERIAL AND THEREFORE THE IMAGE QUALITY WAS POOR. THERE WAS NO PATIENT INVOLVEMENT.