inforMED
MalfunctionQJZ

XLUNG KIT 230

Received May 28, 2026 · Event occurred May 12, 2026

Report 3012172416-2026-00041 · MDR key 25314756

Device

Generic name

Extracorporeal System For Long-term Respiratory / Cardiopulmonary Failure

Manufacturer

Xenios Ag

Catalog number

32000014

Lot number

GSXH2504

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Output Problem
  • Insufficient Device Problem Information

Patient

Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.

Description of Event or Problem

A USER FACILITY REPORTED THAT A PATIENT WAS RESUSCITATED DURING PERCUTANEOUS CORONARY INTERVENTION. EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) SUPPORT WAS STARTED ON (B)(6) 2026 AT 13:00 (LOCAL). ON (B)(6) 2026 AROUND NOON, THE PERFORMANCE OF THE OXYGENATOR WAS DEEMED LOW. AT 4.0 L OXYGEN GAS FLOW, 4.5 L OF BLOOD FLOW AND A FRACTION OF INSPIRED OXYGEN OF 100%, THE ARTERIAL OXYGEN WAS AT 14 KPA (PER LOCAL (B)(6) HOSPITAL THAT UTILIZES KPA, EQUIVALENT IS 105 MMHG). ON (B)(6) 2026, THE KIT WAS EXCHANGED. THE BATCH NUMBER OF THE KIT IS UNKNOWN, AS THE PACKAGING HAS BEEN DISCARDED. THE BATCH NUMBER OF THE OXY IS 32190008. THE LOCAL TECHNICAL TEAM HAS NOT DOWNLOADED THE LOGFILE YET BECAUSE THE PATIENT IS CRITICALLY ILL AND BEING CONNECTED TO VENOARTERIAL-VENOUS ECMO SUPPORT AS WELL AS AN IMPELLA. THE MACHINE LOGFILE WILL BE RETRIEVED ONCE THE CONSOLE IS AVAILABLE. THERE WAS NO SPECIFIED PATIENT SERIOUS INJURY. THE COMPLAINT SAMPLE WAS AVAILABLE FOR PRODUCT EVALUATION.