inforMED
MalfunctionFPK

PRESSURE MONITOR TUBING

Received May 27, 2026 · Event occurred Apr 28, 2026

Report 3011642792-2026-00076 · MDR key 25306161

Device

Generic name

Tubing, Fluid Delivery

Catalog number

PM6148/CNB

Lot number

I3211950

Product problems

  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SUSPECT DEVICE WAS NOT RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. THE ROOT CAUSE COULD NOT BE DETERMINED. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND THREE SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE FOUND. THE PRODUCT HISTORY WAS REVIEWED, AND NO EXCEPTION DOCUMENTS WERE FOUND. NONCONFORMANCES OF THIS NATURE ARE MONITORED BY THE OPERATION TEAM. THIS COMPLAINT WILL BE USED TO TREND FOR SIMILAR COMPLAINTS.

Description of Event or Problem

IT WAS REPORTED THAT DURING SURGERY, A HIGH-PRESSURE SYRINGE ANGIOGRAPHY WAS PERFORMED, AND CONTRAST AGENT OVERFLOW WAS OBSERVED AT THE CONNECTION OF THE PRESSURE EXTENSION TUBE. UPON INSPECTION, THE CONNECTION WAS FOUND TO BE BROKEN. AFTER REPLACING THE PRESSURE EXTENSION TUBE AND CONFIRMING IT WAS CORRECT, ANOTHER BREAK OCCURRED DURING INSTALLATION AND INJECTION. SUBSEQUENTLY, BOTH THE HIGH-PRESSURE SYRINGE BARREL AND THE EXTENSION TUBE WERE REPLACED, AND NO FURTHER ABNORMALITIES OCCURRED. THE SURGERY WAS COMPLETED SUCCESSFULLY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.