inforMED
MalfunctionFLE

EVOLUTION TRANSFER CARRIAGE

Received May 27, 2026 · Event occurred Apr 29, 2026

Report 3005899764-2026-00040 · MDR key 25302607

Device

Generic name

Transfer Carriage

Model number

AX083020003

Catalog number

AX083020003

Product problems

  • Unintended Movement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE UNIT AND FOUND THAT THE LOCKING WHEELS WERE NOT PROPERLY ADJUSTED CAUSING THEM TO DISENGAGE AND REMAIN IN THE EXTENDED POSITION WHEN THE CARRIAGE WAS UNDOCKED. THIS ALLOWED THE RAILS TO SLIDE INSTEAD OF THE WHEELS ROLLING, CAUSING THE REPORTED EVENT. TO RESOLVE THE ISSUE, THE TECHNICIAN ADJUSTED THE LOCKING MECHANISM OF THE TRANSFER CARRIAGE, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. THE TECHNICIAN REMOVED THE DAMAGED LOADING CART FROM SERVICE AND IS PENDING A REPLACEMENT. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE LOADING CART IS REPLACED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER, THEIR EVOLUTION TRANSFER CARRIAGE BECAME UNSTABLE AND FELL, DAMAGING THE LOADING CART. NO REPORT OF INJURY.